The LD Series Regulated DC Power Supply Is A Dual-output Digital Laboratory Power Supply Range Designed For Applications Requiring Stable, Accurately Regulated DC Voltage And Current. With Power Configurations Ranging From 300W To 1280W, The Series Provides Multiple Voltage And Current Combinations For Laboratory Testing, Electronics Development, Servicing, Production And Other Demanding Applications.
The Series Combines phase-controlled Preregulation With Linear Post-regulation To Deliver High Stability And Close Output Regulation. Its Carefully Controlled Output Makes It Suitable For Equipment And Circuits That Require Dependable DC Power During Development, Testing And Measurement.
The LD Series Offers Several Output Configurations, Allowing Users To Select A Power Supply According To Their Required Voltage And Current Range. Available Models Include Versions Providing Outputs Such As 0–32V, 0–64V And 0–128V, With Current Ratings Extending Up To 20A Depending On The Model.
The Range Includes Models Such As LD3205, LD3210, LD3220, LD6405, LD6410, LD1282 And LD1285, Providing Flexible Combinations For Different Laboratory And Electronic Testing Requirements.
The Power Supply Is Designed For Close Voltage And Current Regulation. Both Constant-voltage And Constant-current Operating Modes Are Supported, Helping Maintain Controlled Output Conditions When The Connected Load Changes.
The Documented Regulation Specifications Include CV ±0.01% + 2mV And CC ±0.1% + 2mA. This Level Of Regulation Makes The Series Appropriate For Precision Laboratory Work, Circuit Development And Controlled DC Power Applications.
The LD Series Provides A Maximum Output Ripple Specification Of 1mV RMS, Helping Maintain A Clean DC Output For Sensitive Electronic Circuits And Measurement Applications. Reduced Ripple Can Be Particularly Valuable When Powering Test Setups, Instrumentation And Electronic Assemblies Where Unwanted AC Components Can Affect Measurements Or Circuit Behavior.
Remote Sensing Is Provided To Compensate For Voltage Drops Between The Power Supply And The Connected Load. This Feature Helps Maintain The Required Voltage At The Load Terminals, Particularly Where Longer Output Leads Or Higher Current Operation Can Introduce Measurable Cable Losses.
The Series Features 3-digit Seven-segment Displays For Monitoring Voltage And Current. A Three-digit Digital Panel Meter Provides A Straightforward Way To Observe Operating Values During Testing And Adjustment.
Meter Accuracy Is Specified As ±3 Counts, While Front-panel Controls Allow Users To Set Voltage And Current Conveniently.
Front-panel Potentiometers Provide Separate Voltage And Current Adjustment, With coarse And Fine Controls Available For Precise Setting. The Single-turn Potentiometer Arrangement Makes It Easy To Establish The Desired Operating Point For Laboratory Testing And Electronic Development.
The LD Series Incorporates Protection Functions Designed To Help Safeguard The Power Supply And Connected Equipment. Overload And Short-circuit Protection Are Provided Through A Constant-current Type Arrangement, While Overvoltage Protection Is Available As An Optional Feature.
An Input MCB Provides Convenient Input Switching And Protection.
The Power Supply Has A Compact 3U-high, 19-inch Rack-adaptable Design, Making It Suitable For Laboratory Benches As Well As Equipment Racks And Test Systems. The Approximate Chassis Dimensions Are 430mm × 133mm × 450mm, Excluding Legs And Accounting For The Specified Dimensional Convention In The Source Document.
For Applications Requiring External System Integration, The LD Series Can Be Configured With analogue Programming And Monitoring For Voltage And Current As An Optional Feature. A 19-inch Rack-mounting Arrangement Is Also Available As An Optional Configuration.
The LD Series Can Be Used For A Wide Range Of Regulated DC Power Requirements, Including:
LD Series
Regulated DC Power Supply
Dual Output Digital Lab Series
300W to 1280W
230VAC ±10%
50Hz
1 Phase
LD3205, LD3210, LD3220, LD6405, LD6410, LD1282, LD1285
0–32V / 0–64V / 0–128V, model dependent
Up to 20A, model dependent
0–32V, 0–5A
0–32V, 0–10A
0–32V, 0–20A
0–64V, 0–5A
0–64V, 0–10A
0–128V, 0–2.5A
0–128V, 0–5A
±0.01% + 2mV
±0.1% + 2mA
±0.01% + 2mV
±0.1% + 2mA
1mV RMS maximum
1mV RMS maximum
Provided
0–50°C
Overload / short circuit, constant-current type
Optional
CV / CC
3-digit DPM
Voltage and current
±3 counts
MCB
V set & I set
Single-turn potentiometers
Coarse and fine
19-inch rack adaptable, optional
Optional for voltage and current
3U
Approximately 430W × 133H × 450D mm
Model dependent; approximately 20–30kg for listed configurations
Phase-controlled preregulation with linear post-regulation
High current density, remote sensing and close regulation